JON P. McCALLA, Chief Judge.
Defendant alleges that Plaintiffs' claims are preempted by federal law. This issue is governed by the Supreme Court cases of Riegel v. Medtronic, Buckman v. Plaintiffs' Legal Committee, and Medtronic v. Lohr,
To the extent that Plaintiffs' claims are based on the following regulatory violations, they are impliedly preempted:
Claims premised on Plaintiffs' derivative assertion that the HiRes90K device with the AstroSeal feedthru was "adulterated" or "misbranded" pursuant to 21 U.S.C. § 351 are also preempted.
To the extent that Plaintiffs' claims are premised on violation of, or deviation from, the following requirements, the claims are not preempted:
To the extent that Plaintiffs claims are premised on alleged violations of the other Current Good Manufacturing Practice ("CGMP") regulations cited by Plaintiffs, Plaintiffs have not stated actionable parallel claims. The CGMP regulations cited are too generic to provide a requirement that could support a parallel claim.
Under Riegel, Kemp, and Howard, claims premised on the following conduct may support a parallel claim, and thus survive express preemption under 21 U.S.C. § 360k(a): (1) a violation of FDA regulations; (2) a violation of, or deviation from, the FDA requirements established for a Class III device through the PMA process; and (3) a violation of, or deviation from, a specific and enforceable CGMP regulation.
Under Buckman, Kemp, and Cupek, private rights of action to enforce FDA administrative and reporting requirements are prohibited by 21 U.S.C. § 337(a). Thus, even if a claim survives express preemption, it may be impliedly preempted if it amounts to a disguised fraud-on-the-FDA claim.
To the extent that Plaintiffs' claims are based on Advanced Bionics' purported failure to submit a PMA Supplement Application and obtain PMA approval pursuant to 21 C.F.R. § 814.39 for the HiRes90K with the AstroSeal feedthru, the claims are impliedly preempted. PMA approval is an administrative requirement, not a substantive safety requirement. Thus, claims premised on PMA approval are disguised fraud-on-the-FDA claims and, therefore, impliedly preempted.
Likewise, claims premised on Plaintiffs' derivative assertion that the HiRes90K device with the AstroSeal feedthru was "adulterated" or "misbranded" pursuant to 21 U.S.C. § 351 are also preempted.
To the extent that Plaintiffs' claims are based on Advanced Bionics' purported failure to identify the change to the AstroSeal feedthru in its annual report pursuant to 21 C.F.R. § 814.84, the claims are impliedly preempted. Annual reporting requirements are administrative requirements, not substantive safety requirements. Thus, claims premised on reporting requirements are disguised fraud-on-the-FDA claims and, therefore, impliedly preempted.
To the extent that Plaintiffs' claims are premised on the allegation that Advanced Bionics' deviated from the manufacturing and design requirements set forth in PMA Supplement 30 for the HiRes90K by changing from the PA & E feedthru to the AstroSeal feedthru, the claims are not preempted.
In Kemp, the Sixth Circuit explained that:
Advanced Bionics' admits, for purposes of its preemption motion only, that it did not submit a PMA Supplement for the HiRes90K with the AstroSeal feedthru. Plaintiffs have offered proof that the feedthru assembly is a critical component, and thus, the change to the AstroSeal feedthru was a change that affected the safety and effectiveness of the HiRes90K. Accordingly, Plaintiffs' claim that Advanced Bionics deviated from its PMAS 30 requirements when it modified the HiRes90K and changed to the AstroSeal feedthru is a parallel claim that survives preemption.
To the extent that Plaintiffs' claims are premised on the allegation that Advanced Bionics did not test finished HiRes90K devices with AstroSeal feedthrus under actual or simulated use conditions in a violation of 21 C.F.R. § 820.30(g), the claims are not preempted.
Section 820.30(g) is a CGMP regulation that is part of the umbrella Quality System Regulations applicable to all medical devices under 21 C.F.R. Part 820. Pursuant to 21 C.F.R. § 820.30(g), a manufacturer is required to validate the device design, which "shall include testing of production units under actual or simulated use conditions." In Howard, the Sixth Circuit made it clear that CGMP regulations may impose
Thus, Advanced Bionics' argument that this claim is expressly preempted because it performed all the testing outlined in PMA Supplement 30 for the HiRes90K is unavailing. Section 820.30(g) provides that a device must be tested in an environment that simulates the end use environment. This testing requirement is imposed on all manufacturers of Class III devices. Thus, it does not impose requirements that are "different from" or "in addition to" those prescribed by the FDA.
Plaintiffs also premise their claims on alleged violations of several other CGMP regulations: 21 C.F.R. §§ 820.20, 820.80, 820.100, and 820.198. These CGMP regulations are too generic to provide a requirement that could support a parallel claim.
In holding that the particular CGMP regulation at issue could support a negligence per se claim, the Howard court was careful to distinguish In re Medtronic, Inc. Sprint Fidelis Leads Prods. Liab. Litig.
This Court reads Howard as distinguishing between CGMP regulations that are "so vague as to be incapable of enforcement," and those that impose a concrete requirement on a manufacturer that embodies a standard of care related to the safety and effectiveness of the device. In this regard, the question of whether a CGMP regulation may support a parallel claim is similar to the question of whether the violation of a statute can support a negligence per se claim. If the CGMP regulation in question sets forth a specific and substantive standard of care that is intended to protect others, then it imposes a requirement on the manufacturer. As a result, the violation of that CGMP regulation may support a parallel claim and, incidentally, a negligence per se claim.
On the other hand, CGMP regulations that amount to nothing more than administrative requirements or generalized objectives impose no duty on the manufacturer and, thus, may not support a parallel claim.